ZTF Pressure Gauge
Pfieffer Documentation
Configuration Instructions
The following configuration file should be copied to a USB memory stick:
- SETUP02.CSV: setup file to configure TPG361 on ZTF/P48 network (IP 192.168.1.27)
The configure the pressure gauge controller with this file using the following procedure:
- Power off controller
- Insert memory stick
- Power on controller
- Select set up
- Restore set up
- Select #2
- Power off controller
- Unplug memory stick
- Start recording pressure by holding down the up arrow, or use the tpg.py script.
When Logging Fails
If the pressure isn't being logged, run monitord_test, there is an option there to turn the Pfeiffer on and off. Watch the monitord.log file to see when it connects and starts reading properly again.
--
DavidHale - 03 Nov 2017